DF-310E
Flexible, Adaptable Oxygen Analyzer

The DF-310E with the E-Sensor is ready to
handle your oxygen analysis needs in almost any application.
The DF-310E features:
- Accuracy the greater of ±3 reading
or ±0.02% range
- Ranges from 0-0.5ppm to 25%Instantaneous response
to oxygen change
- Fast response typically less than 10 seconds
to read 90% of a step change
- Background gas compatibility including
N2, H2, Co, freons, hydrocarbons, and more
- STAB-EL option removes
acids and ionic impurities from the electrolyte that could affect
sensor performance
- Optional battery back-up for extended use and
process protection
Range
Ranges are available from 0-50 ppm to 0-25%
Sample Gas Compatibility
BASIC SENSOR: All inert and passive gases including N2, CO, H2,
Ar, freons, hydrocarbons, etc.
SENSOR WITH Stab-El OPTION: Neutralizes trace contaminants
including acids such as CO2, H2S, NOX, SOX, etc. (Consult Delta
F for concentration limits.)
Options
Stab-ElAcid Gas System
Sample Pump
Oxygen Alarms
Flow Alarm
Isolated 4-20 and 2-20 mADC outputs
NEMA 4X and 7 Sensor Enclosures
Remote Sensor Heater
Battery Backup
Accuracy (at
constant temperature)
Standard Resolution Sensor: Greater of ±3% of reading or
0.5% of range
High Resolution Sensor: Greater of ±3% of reading or 0.02%
of range (except for ranges <100 ppm: ±3% of reading or
±0.05% of range)
Response Time
Responds instantaneously to O2 change. Equilibrium
depends on specific conditions. Typically <10 seconds to read
90% of a step change.Sample Requirements
Gas phase, non-condensing, 0° to 50° C (32° to 122°
F)
Flow from 1.0 to 3.0 scfh at 0.2 to 1.0 psig pressure (except 15-25
psig for 310E H0050 without pump)
Misc.
21.1cm W x 19.7cm H x 20.1cm D 4.52 kg. (8.32"W
x 7.75"H x 7.91"D 10 lbs.)
Operate between 0° and 45° C ambient (32° and 120°
F)
Input Power: 24 VDC, 1 Amp (max)
For more information, click here to view the
DF-300E
Series brochure.
Download the print
resolution DF-300E Series brochure.
For more information, click here to view
the DF-310E Datasheet pdf.
View the DF-310E
Specification & Configuration sheet.
Open the E-Sensor
Technical Bulletin
Download the print
resolution E-Sensor Technical Bulletin
Download Recommendations: Because of occasional
browser conflicts, and load times we recommend that you download
print resolution pdf files to your system by right-clicking on the
link for PC users - click and hold for Macintosh users. |