DF-150E
Custom Configure the Exact Oxygen Analyzer
You Need

The DF-150 features:
- Maximum configuration flexibility build
it the way you need it
- Superior performance out-of-the-box
- Accurate - ±2% full scale at constant
temperature
- Instantaneous response to oxygen change
- Fast response typically less than 10 seconds
to read 90% of a step change
Every DF-150E is custom configured under ISO-9001
controls to your precise specifications and calibrated to standards
traceable to NIST and operated for several days to ensure flawless
performance. The ultra-stable E-Sensor can operate for years without
re-calibration giving you a long and useful life.
DF-150E Specifications
Range
Ranges are available from 0-10 ppm to 0-25%
Sample Gas Compatibility
STANDARD SENSOR: All inert and passive gases including N2, CO, H2,
Ar, freons, hydrocarbons, etc.
SENSOR WITH Stab-El OPTION: Neutralizes trace contaminants
including acids such as CO2, H2S, NOX, SOX, etc. (Consult Delta
F for concentration limits.)
Options
Oxygen Alarms
Stab-El Acid Gas System
CE Certification
Analog and Digital Outputs
Classified Area Enclosures
Flow Alarm
Three range
Sample Pump
Remote Sensor
Accuracy (at
constant temperature)
Greater of ±3% of reading or 0.5% of range
Response Time
Responds instantaneously to O2 change.
Equilibrium depends on specific conditions. Typically <10
seconds to read 90% of a step change.Sample Requirements
Gas phase, non-condensing, 0° to 150° F.
Flow should be 1.0 to 3.0 scfh at 0.2 to 1.0 psig pressure.
Misc.
19.7cm W x 21.6cm H x 25.4cm D
7kg. (7.75"W x 8.5"H x 10"D 15
lbs.)
Operate between 0° and 45° C ambient (32° and
120° F)
115 VAC or 220 VAC at 25 Watts
For more information, click here to view
the DF-150E Datasheet pdf.
View the DF-150E Specifications & Configuration Sheet
|