DF-140E
Reliable, Accurate Oxygen Sensor in a Nema
4x Enclosure

Superior Oxygen Analysis for Harsh and Hazardous
Environments
The DF-140E features:
- Long term reliability and stability
- Durability can be used in Class
I, Division 1 or 2 areas
- Low/No maintenance
- STAB-EL option allows for accurate
measurement in the presence of acid gases
The DF-140E allows for reliable oxygen measurement
in a wide variety of environments, including outdoors and in explosive
environments with a NEMA 7 remote sensor enclosure. Using the revolutionary
non-depleting E-Sensor, the DF-140E delivers reliable readings without
frequent recalibration and periodic sensor replacement.
DF-140E Specifications
Range
Ranges are available from 0-10 ppm to 0-25%Sample
Gas Compatibility
STANDARD SENSOR: All inert and passive gases including N2, CO, H2,
Ar, freons, hydrocarbons, etc.
SENSOR WITH Stab-El OPTION: Neutralizes trace contaminants
including acids such as CO2, H2S, NOX, SOX, etc. (Consult Delta
F for concentration limits.)
Options
CE Certification
Stab-El Acid Gas System
Oxygen Alarms
Sample Pump
Flow Alarm
Three range
Isolated 4-20 mADC output
Remote Sensor Heater
Accuracy (at
constant temperature)
Greater of ±3% of reading or 0.5% of range
Response Time
Responds instantaneously to O2 change.
Equilibrium depends on specific conditions. Typically <10
seconds to read 90% of a step change.
Sample Requirements
Gas phase, non-condensing, 0° to 50°
C (32° to 122°F)
Flow should be 1.0 to 3.0 scfh at 0.2 to 1.0 psig pressure.
Misc.
35.6 cm W x 33.5cm H x 29.0cm D
13.6kg. (14.0"W x 13.18"H x 11.42"D
30 lbs.)
Operate between 0° and 45° C ambient (32° and
120° F)
115 VAC or 220 VAC at 25 Watts
For more information, click here to view
the DF-140E Datasheet pdf.
View the DF-140E Specifications & Configuration Sheet |